+7(962)-699-12-53
0
Корзина
0
Товар добавлен в корзину!
Каталог товаров
0
Избранные
Товар добавлен в список избранных

Endomed 582 User Manual ((link)) Review

Connect the device to a properly grounded AC outlet.

Independent rotary knobs for precise current adjustment per channel. 2. Operational Interface and Control Panel

Connect the electrode cables to the channel outputs, ensuring red plugs go to red sockets and black to black. endomed 582 user manual

Disposable pads for high-hygiene environments

Do not apply current over deep vein thrombosis (DVT) or active thrombophlebitis. Connect the device to a properly grounded AC outlet

| Error Display | Meaning | Action | |---------------|---------|--------| | | Electrode cable not connected or broken electrode lead. | Check plug connections. Test cable continuity with a multimeter. | | "Output High" | Intensity exceeds the safe limit for current path (e.g., electrodes too close together). | Reduce intensity immediately. Reposition electrodes. | | Internal Beeping + Blank Screen | Thermal overload or internal fuse failure. | Power off for 15 minutes. If persists, return for service. | | "E-01" | Microprocessor communication error. | Cycle power. If repeated, factory reset required. |

To ensure the functions correctly, regular maintenance is required. | Check plug connections

Thoroughly rinse sponges under running water and allow them to air dry completely. Replace them regularly to prevent bacterial growth. Troubleshooting Common Issues Potential Cause Recommended Solution Device will not turn on Disconnected power cord; Blown fuse.

Clean reusable rubber electrodes thoroughly after each use with warm water and soap to remove residual conductive gel.

Check the timer settings or the electrode contact quality. Accessing the Full Endomed 582 Manual

0
Избранные
Товар добавлен в список избранных
0
Корзина
0
Товар добавлен в корзину!
×